value (Kt or Kt/V) is adjusted after every successfully
performed measurement check throughout the treatment.
Time To Target
If a target value for Kt/V (Kt/V TARGET) has been
set, the machine will show an estimate of the treatment
time needed to reach the desired Kt/V target value.
Ti me to target will be calculated by the machine and
will automatically be shown as soon as the rst Diascan
measurement check has been performed. This estimated
time can be compared with the treatment time set but
there will not be an attention alarm generated to make the
operator attentive to any differences.