Profiling Models
The proling function of the AK 96 dialysis machine consists of three
different models (memories). Proles for sodium (NA), bicarbonate
(HCO3) and ultraltration rate (UF) can be set for each model. It is
possible t o set proles for all three parameters at the same time or
just pick one or two parameters within the model. The models can be
manually set or they can be preset by an authorized technician.
Profiling Setting/Activation
Proling can be started at treatment start or at any time during
treatment. The model that is to be used can be manually set or be preset
by an authorised technician.
Step-by step instructions for the two alternatives are described further
on in this section, in "Pro ling without Preset Model", page 7:7 or in
"Proling with Preset Model", page 7:10 in part 1.
Proling settings can be done as soon as the machine has been
switched on and “FCh” is shown on the Time Display. It is possible
to manually set a model or just select a preset model. Proles for
sodium (NA), bicarbonate (HCO3) and ultraltration rate (UF) can be
set/adjusted for each model. If UF proling (UF)istobeused,the
UF volume for the following treatment has to be set before setting
the UF proling.
The proling function for all three proling parameters can be reached
by rst pressing the Fluid Path button, and then selecting UF or
COND using the Keypad.
Activation The proling function can be activated at the same time as
when the setting is being done, or at any other time desired during
treatment. When UF proling has been activated it is possible to
change the proling settings without rst deactivating proling.
However, changing sodium and/or bicarbonate proling settings cannot
be done without rst deactivating proling. The proling function
starts to be active when blood has been detected in the venous line and
the UF Start/Stop button has been pressed.
7:6
AK 96
®
Operator's Manual - Profiling
HCEN12239 Revision 12.2010
Program version 3.xx

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gambro AK 96 and is the answer not in the manual?

Gambro AK 96 Specifications

General IconGeneral
BrandGambro
ModelAK 96
CategoryMedical Equipment
LanguageEnglish

Related product manuals