FLUID TUBES IN SAFETY BYPASS
Connect fluid tubes to dialyzer
Appears:
This attention alarm can appear for three different
reasons:
1. If the dialysis fluid tubes are still connected to the
safety couplings and blood is detected by the priming
detector.
2. If the dialysis fluid tubes are reconnected to the safety
couplings during treatment. Buzzer is activated 5
min. after attention alarm is generated.
3. During assisted priming if the dialysis fluid tubes are
still connected to the s afety couplings and they are
expected to be connected to the dialyzer.
Machine actions:
None
Measures:
Move the dialysis fluid tubes from the safety couplings
to the dialyzer.