UF VOLUME SET TO ZERO
Press SELECT key to confirm
Appears:
When the priming detector detects blood and the set UF
volume is zero.
Machine actions:
Buzzer is activated 2 min. after attention alarm is
generated.
Measures:
Press the Select key to conrm no ultraltration during
treatment.
Mute tim e:
2min.
Buzzer sound:
Soft buzzer
1
4:128
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order