Procedure for manual activation
1. Press the Fluid path button, select DIA FLUID FLOW using
the Keypad.
2. Press the Select key for 3 seconds to activate concentrate
stand-by mode.
3. The CONCENTRATE STANDBY MODE menu will now be
displayed on the Information Display.
Resume fluid preparation
To resume the dialysis fluid preparation (Concentrate stand-by
mode deactivated) press the Select key for 3 seconds when the
CONCENTRATE STANDBY MODE is displayed. The dialysis fluid
will be ready within approximately 2 minutes.
If dialysis fluid preparation has not been resumed during priming, it
will automatically be resumed when blood is detected in the priming
detector after treatment start. Dialysis fluid preparation will also
automatically be resumed when concentrate stand-by mode has been
active for more than one hour.