Residual Test after Chemical Disinfection
Before connecting to a patient, a residual test must be done on the
prepared dialysis fluid. This is to make sure that the patient will not be
exposed to chemical residuals after a chemical disinfection program.
The residual concentration in the dialysis fluid must be b elow levels
specified by the clinic or by national standards. It is essential to use an
appropriate test method, either with proven sensitivity for the chemical
or recommended by the manufacturer of the chemical disinfectant.
CAUTION
After the Chemical disinfection procedure a test for residues must
have been performed prior to connecting to a patient.
CAUTION
Note
• The pick-up tubes must be rinsed/disinfected separately. See
"Pick-up Tubes” on page 8:41 for instructions.
Note
Procedure
1. Switch the machine on and let it pass function check with the
proper concentrates attached.
2. When the green fluid path lights up, press the fl
ashing
Fluid bypass button in order to let fresh dialysis fluid flow
through the dialysis fluid tubes.
3. Take the residual test sample from t
he flowing fluid on the
dialysis fluid outlet tube (to dialyzer).
If the test shows disinfectant r esidues:
• DO NOT u se connected BiCart
cartridge.
• Press the Rinse/Disinfection button and start a new rinse/drain
procedure (the button lights up). When this is finished, r epeat
the handling procedure
described above.
If the machine still shows disinfectant residues, contact an authorized
technician.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Hygiene and Maintenance
8:31