10. If a new treatment is to be started, press the flashing
Priming button to initiate a new function check instead of
switching off the machine.
11. Switch off the machine by pressing the On/Off button for
3 seconds.
12. If Formal 37% program has been performed, the ultrafilter
used in UFD must be changed prior to the following
treatment. Alternatively perform additional rinse/drain
procedures according to section “Residual Test after
Chemical Disinfection", page 8:31.
The machine can be preset to automatically switch off a fter a complete
chemical disinfection program, i.e. when the Priming button is
flashing, by an authorized technician in the clinic.
CAUTION
After the Chemical disinfection procedure a test for residues must
have been performed prior to connecting to a patient.
CAUTION
Note
• The pick-up tubes must be rinsed/disinfected separately. See
"Pick-up Tubes” on page 8:41 for instructions.
Note
8:30
AK 96
®
Operator's Manual - Hygiene and Maintenance
HCEN12239 Revision 12.2010
Program version 3.xx