Ultrafilter - How to change
Note
Make sure that the ultralterischangedinanasepticway.
Note
Procedure
1. Release the lower part of the ultralter holder by pulling t he handle
and pressing the latch downwards.
2. Remove the ultralter by pulling it gently downwards.
3. Insert the new ultralter into the holder and push it gently upwards.
4. Close the holder by pushing the lower latch into position.
5. Conrm the change of ultralter (see Ultralter replacement
reminder on page 8:39).
6. Label the ultra lter with date for change.
7. Perform a Heat disinfection program before the AK 96 dialysis
machine is used for treatment after the ultralter has been changed.
Ultrafilter replacement reminder
If UFD (Ultralter Dialysis Fluid) is inst
alled the operator will
be reminded to change ultralter. The operator will be notied
by an attention alarm when a limit has been exceeded. There are
three different limits: number of
days since last change, number
of heat disinfections since last change and number of hypochlorite
disinfections since last change.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Hygiene and Maintenance