To remove the blood pump rotor; hold the handle and pull
out. Then turn clockwise until the blood pump rotor loosens
from the pump shaft.
To attach the blood pump rotor; whilst holding the handle,
place the rotor on the pump shaft. Then move your fingers
from the handle and place them on the rotor as shown in the
figure. Turn the r otor clockwise slowly, and at the same time
push slightly, until it reaches the bottom position. Continue
turning until the blood pump handle clicks in.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Machine Description
2:9