EasyManuals Logo

Gambro AK 96 User Manual

Gambro AK 96
509 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #142 background imageLoading...
Page #142 background image
2. Select alternative in the
ASSISTED PRIMING SETUP MENU using the Keypad.
3. Press the Select key and the menu will change.
If the d ialysis uid preparation is complete ("F.Ch" disappears
from the Time Display and the bypass path on the Flow Diagram
lights up) the dialysis uid tubes can be connected to the dialyzer.
See "Dialysis Fluid Tubes - Attach", page 4:35 in part 1 for
instructions on how to attach the tubes. Especially note that if
UFD (dialysis uid lter) is installed, the dialysis uid tubes
cannot be moved until the bypass path of the Flow Diagram
lights up green.
If the dialysis uid preparation is not complete, the process
can be started with the dialysis uid tubes connected to the safety
couplings.
4. Press t he ashing Blood pump button to start t he blood pump
and the procedure. If the blood pump segment has been
automatically threaded, check for possible kink
s. Note that
the dialyzer should be lled from bottom to top to ensure
proper de-airation.
The procedure starts with the ll up pha
se.
ACTIVE PHASE:FILLUP will be displayed.
5. Fill up the level in the venous
drip chamber (while the
blood pump is running) using the level adjustment knob
(anticlockwise direction). Adjust the level so it is well above
the air detector head.
6. The Airdetectorbuttonwill start to ash when the
venous drip chamber has been lled. This is to r equest
the operator to pres
s it and by doing so, activate the air
detector alarm function. It is not necessary to press the
ashing button at this moment, the air detector activation
may be moved for
ward until the extracorporeal circuit is
completely de-aired. Additionally, if the air detector is not
manually activated during priming, it will be automatically
activated ei
ther when the priming volume is achieved, when
CONNECT PATIENT is selected or when blood is detected.
The activated air detector alarm function is indicated by an
unlit but
ton.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:47

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gambro AK 96 and is the answer not in the manual?

Gambro AK 96 Specifications

General IconGeneral
BrandGambro
ModelAK 96
CategoryMedical Equipment
LanguageEnglish

Related product manuals