Invasive blood pressure
14-7
Table 14-1 Invasive blood pressure labels and descriptions
Cerebral perfusion pressure
Cerebral Perfusion pressure CPP is calculated by subtracting ICP mean pressure from Art mean
pressure.
CPP mean is displayed next to the ICP value.
NOTE: Set the Label of the channel to ICP, and the Digit Format to CPP.
Adjusting alarm sources and limits
You can adjust or turn off pressure alarm limits in the Alarms Setup/Adjust Limits menu. Enter
the menu by pressing the
Alarms Setup key.
For each pressure channel you can choose as a source Sys (Systolic), Dia (Diastolic), Mean or
Off. Note that you can choose one or both sources to be active at a time.
For more information about alarms and adjusting, see section ”Alarms.”
LABEL P1, Art,
ABP
P2, CVP RAP, LAP ICP PA RVP
Scale mmHg/kPa 200/30 20/3 20/3 20/3 60/8 60/8
Color Red Blue White White Yellow White
Alarm source Sys Mean Off Off Off Off
Digit format S/D Mean Mean CPP S/D S/D
Filter (Hz) 2299999
Response normal normal normal normal normal normal
Alarms
Setup