Alarms
4-11
Deactivating alarms
You can set the alarm limits temporarily to OFF for the following parameters: HR, SpO
2
and
respiration rate. This way, you can silence a parameter without having to adjust the alarm
limits.
1. Press the
Alarms Setup key.
2. Select
Adjust Limits.
3. Turn the ComWheel to highlight the measurement.
4. and push the ComWheel to change the selection
Off.
The symbol appears in the digit field.
When you select the alarm limit ON again, the previous alarm limits will be active.
Automatic recording on alarms
An automatic local recording is possible when Asystole, Tachy/Brady, Art High/Low, V Fib or V
Tachy alarms reach the red alarm level. When the red alarm level is reached, the recorder
prints the ECG + Art waveforms.
1. Press the
Print/Record key.
2. Select Record Waveforms.
3. Select Start on Alarms and Yes.
Print/
Record