EasyManuals Logo

GE Healthcare B30 User Manual

GE Healthcare B30
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #56 background imageLoading...
Page #56 background image
B30 Patient Monitor
2-30
Performance
WARNING Operation of the monitor outside the specified values may cause inaccurate
results.
Any fluctuations within the specified limits do not affect the performance.
NOTE: Information in this section can be especially useful to clinicians.
Power supply
Rated voltages and frequencies: 100 to 240 V 50/60 Hz
Allowed voltage fluctuations: ±10 %
Maximum power consumption: 150 VA
Battery operation
Batteries: Exchangeable lithium-ion, 2 pcs max.
Charging time: 2 hours per battery pack
Operation time: up to 4.5 hours
Environmental conditions
Operating temperature:
normal operation: +5 to +40°C (41 to 104°F)
while charging batteries: +5 to +35°C (41 to 95°F)
Storage and transport temperature: -20 to +60°C (-4 to 140°F)
Relative humidity: 10 to 90 % noncondensing
Atmospheric pressure: 670 to 1060 mbar (500 to 800 mmHg)
Alarm behavior
The maximum alarm delay of the alarm at the monitor signal output to network: 5 seconds
If the alarm mode is latched, the technical alarms are latched as well. This does not comply with the NIBP
(IEC 60601-2-30) and invasive pressure (IEC 60601-2-34) standard requirements.
Silencing alarms for 5 minutes does not comply with the SpO
2
(ISO 9919) standard requirements.
Defibrillator & IABP synchronization connector
Analog output
ECG:
From first user lead (ECG1)
Gain: 1 V/mV ±10%
Delay: < 15 ms
DC offset: ±100 mV max.
Frequency response: 0.05 Hz to 40 Hz
Invasive blood pressure:
From pressure labeled ‘Art’
Gain: 10 mV/mmHg ±2%
Delay: < 35 ms
DC offset: ±20 mV max.
Frequency response: DC to 30 Hz
The pacemaker pulses have been replaced with 2 ms ±20% fixed digital pulses at the ECG analog output.
A device that fulfils the requirements of the IEC 60601-1 standard can be connected to the defibrillator &
IABP synchronization connector. There are no other limitations, because the signals of the connector are
galvanically isolated from patient applied part of the ECG and invasive blood pressure measurements.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Healthcare B30 and is the answer not in the manual?

GE Healthcare B30 Specifications

General IconGeneral
BrandGE
ModelHealthcare B30
CategoryMedical Equipment
LanguageEnglish

Related product manuals