B30 Patient Monitor
2-12
Batteries
The monitor can be run either on mains power or batteries. Battery operation is initiated when
the power cord is disconnected or when the mains power is lost during monitoring.
NOTE: Always use the B30 with batteries inserted. Otherwise all trend data and
temporary settings are lost if the power cable is detached from the mains.
NOTE: Before using the monitor for the first time, charge the batteries to their full
capacity. Charging time is two hours per battery pack.
The B30 has two lithium-ion batteries at most, located in the battery compartment. They can
be charged separately, and screen symbols and monitor frame LEDs indicate their charging
level and possible failure, see below. You can also check the battery status through
Monitor
Setup
- Battery Setup. The internal battery capacity is up to 4.5 hours with fully charged
batteries.
NOTE: When the monitor is battery powered, the green battery LED is on. When the
monitor is mains powered, the green mains LED is on.
If you wish to have the battery charge visible at all times, select it in one of the digit fields:
Monitor Setup - Screen Setup - Digit Fields - Battery. You can now see how much charging
time is left for each battery separately both in numbers and as symbols, and the total charging
time in numbers.
WARNING Use only manufacturer approved batteries. Contact GE service if you need
to order new batteries.
Monitor
Setup

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Healthcare B30 and is the answer not in the manual?

GE Healthcare B30 Specifications

General IconGeneral
BrandGE
ModelHealthcare B30
CategoryMedical Equipment
LanguageEnglish

Related product manuals