2 Mounting Requirements
2.1 Floor Loading and Recommended Mounting Methods
See Table 5-2. To obtain floor loading and recommended mounting methods for components not
specified in Table 5-2, refer to the appropriate component Pre-Installation Manual listed in Chapter
2, Basic Innova LC System Compatibility.
Table 5-2:
PRODUCT OR
COMPONENT
NET WEIGHT
KG (LBS)
DIMENSIONS MM (INCHES)
LOAD BEAR‐
ING AREA
MM (INCHES)
WEIGHT/ OC‐
CUPIED
AREA
MOUNTING METHOD
WIDTH
DEPTH
HEIGHT
Liberty LC Position‐
er
750 (1655)
See Illustrations
INNOVA LC POSI‐
TIONER DIMENSIONS
:
•
Side view
•
Top view
•
Front view
in Dimension Drawings
Circle diame‐
ter 600 (23.62)
Recommended:
•
Through-Bolts
(12)
Alternates:
•
On Grade 5/8 in,
Anchors (12)
•
Above Grade 3/4
in. Anchors (12)
See Illustration 5-28,Il‐
lustration 5-29 and Illus‐
tration 5-30
Table
590 (1300) See
note 1
See illustrations in Dimension Drawings
•
Omega IV
•
Omega V
571.5x429
(22.5x16.9)
2410 kg/m
2
(492.3 lb/ft
2
)
Same as Innova LC Po‐
sitioner See IMPOR‐
TANT NOTICE below.
785 (1731)
Innova
IQ
750x600
(30x24)
2000 kg/m
2
(409 lb/ft
2
IMPORTANT NOTICE
The Table baseplate or the System baseplate are mandatory to install the table (patient support). The Table must never be installed
on grade.
C2 Cabinet
299 (659)
See Illustration
C2 Cabinet Dimen‐
sions
in Dimension Drawings
600x900
(23.63x35.44)
478 kg/m
2
(98
lb/ft
2
)
C1 Cabinet
506 (1115.5)
See Illustration
C1 Cabinet Dimen‐
sions
in Dimension Drawings
600x900
(23.63x35.44)
838 kg/m
2
(172 lb/ft
2
)
COOLIX 4000
204 (450) See
note 3
556 (21.9)
See note 2/3
830 (32.7)
See note 3
1090 (43)
See note 3
Four casters
Not applicable
Detector Chiller
Thermo-Tek
13.6 (30)
See Illustration
Detector Chiller Ther‐
mo–Tek Dimensions
in Dimension
Drawings
283x351
(11.14x13.82)
Detector Chiller
Thermo-Con
14.6 ( 32.2)
See Illustration
Detector Chiller Ther‐
mo–Con Dimensions
in Dimension
Drawings
13.5x11.14
(344x283)
Fluoro UPS UL (op‐
tional)
530 (1169)
690 (27.0)
800 (31.5)
1820 (70.1)
975 kg/m
2
(200 lb/ft
2
)
Ground -mounted
Fluoro UPS CE (op‐
tional)
480 (1059)
680 (26.7)
800 (31.5)
1450 (57.1)
883 kg/m
2
(181 lb/ft
2
)
Ground -mounted
Fluoro UPS IF box
(optional)
4 (9)
280 (11.02)
314 (12.36)
124.5 (4.9)
Wall-mounted
PDB US (4)
148 (326)
685 (27)
225 (9)
1850 (73)
Wall-mounted
PDB CE (4)
110 (242)
800
300
1200
Wall-mounted
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
Chapter 5 Physical Characteristics 139