Center of anchor hole to concrete edge distance 79.4 mm.
Make sure to obtain data on compression strength of the concrete before using floor anchors.
3.3.3 Pan Type Floor Construction Requirement
For Pan type floor construction, steel channels must be designed by a local structural engineer to
span floor joists. See Illustration 3-4.
NOTE: For specific floor preparation procedures, refer to PIM Innova 2000/2100–IQ/
3100/4100 Cardiovascular Imaging System Pre–Installation Kit Installation
Procedures.
Illustration 3-4: Through–Bolt Supplied (Pan Type Floor Construction)
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
68 3 Structural Requirements