20 PIST0021 - US Fluoro UPS and PDB connection
20.1 Personnel Requirements
Personnel
Requirements
Preliminary Reqs Procedure Finalization
1 Not Applicable 1 hour Not Applicable
20.2 Preliminary Requirements
20.2.1 Tools and Test Equipment
Item
Qty Effectivity Part# Manufacturer
Standard Service Engineer’s
toolkit
1 - - -
20.2.2 Required Conditions
Condition
Reference Effectivity
The PDB is already installed - -
Power cables are as follows: (1) From PDB
to UPS rectifier input = 4x6 AWG
recommended (4x3 AWG max) (3ph + PE).
(2) Bypass = 5x8 AWG recommended (5x3
AWG max)(3ph + N+PE). (3) From UPS
output to PDB = 5x8 AWG recommended
(5x3 AWG max)(3ph + N+PE)..
- -
Cable sizing shall follow local requirements. - -
Power cable Mechanical Terminals Torque
Specification: < 3 Nm / < 27 lb.in for 20 AWG
to 3 AWG.
- -
Check the UPS shipping date marked on the
packaging label. Commissioning of the UPS
must be performed within 3 months after
that date.
- -
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
Chapter 6 Electrical Connections 229