Chapter 3 Physical Requirements of Room
1 Presentation of the 3 Rooms
1.1 Examination room
•
Innova LC Positioner
•
Omega / Innova
IQ
Patient Table
•
Performix X–Ray Tube Assembly
•
Collimator
•
Innova Digital Detector
THE ELECTRONIC CABINETS (C1, C2, COOLIX 4000 CHILLER, DETECTOR
CONDITIONER AND FLUORO UPS WHEN INSTALLED) INCLUDE FANS THAT
ARE CREATING AIR–CIRCULATION OF PULSED–AIR. WHEN THIS PULSED
AIR IS IN AN ENVIRONMENT THAT MAY CONTAIN AIRBORNE PATHOGENS
LIKE AN EXAM ROOM/CONTROL ROOM, THERE IS A RISK OF
TRANSMISSION OF THESE AIRBORNE PATHOGENS FROM PATIENTS TO
OTHER PATIENTS OR CLINICAL PERSONNEL (NOSOCOMIAL DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE INSTALLED
IN A ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM, I.E.,
TECHNICAL ROOM.
1.2 Technical room
•
C2 Cabinet
•
C1 Cabinet
•
Coolix 4000 Chiller
•
Detector Conditioner HEAT DRY 1
•
Fluoro UPS (option) (1 Cabinet)
•
Fluoro UPS IF box (option)
1.3 Control room
•
VCIM
•
DL Flat Panel
•
DL Keyboard
•
Monitors