EasyManuals Logo

GE Innova 4100 User Manual

GE Innova 4100
290 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #53 background imageLoading...
Page #53 background image
Chapter 3 Physical Requirements of Room
1 Presentation of the 3 Rooms
1.1 Examination room
•
Innova LC Positioner
•
Omega / Innova
IQ
Patient Table
•
Performix X–Ray Tube Assembly
•
Collimator
•
Innova Digital Detector
THE ELECTRONIC CABINETS (C1, C2, COOLIX 4000 CHILLER, DETECTOR
CONDITIONER AND FLUORO UPS WHEN INSTALLED) INCLUDE FANS THAT
ARE CREATING AIR–CIRCULATION OF PULSED–AIR. WHEN THIS PULSED
AIR IS IN AN ENVIRONMENT THAT MAY CONTAIN AIRBORNE PATHOGENS
LIKE AN EXAM ROOM/CONTROL ROOM, THERE IS A RISK OF
TRANSMISSION OF THESE AIRBORNE PATHOGENS FROM PATIENTS TO
OTHER PATIENTS OR CLINICAL PERSONNEL (NOSOCOMIAL DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE INSTALLED
IN A ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM, I.E.,
TECHNICAL ROOM.
1.2 Technical room
•
C2 Cabinet
•
C1 Cabinet
•
Coolix 4000 Chiller
•
Detector Conditioner HEAT DRY 1
•
Fluoro UPS (option) (1 Cabinet)
•
Fluoro UPS IF box (option)
1.3 Control room
•
VCIM
•
DL Flat Panel
•
DL Keyboard
•
Monitors
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
Chapter 3 Physical Requirements of Room 53

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Innova 4100 and is the answer not in the manual?

GE Innova 4100 Specifications

General IconGeneral
BrandGE
ModelInnova 4100
CategoryMedical Equipment
LanguageEnglish

Related product manuals