2.4.3.3 Recommended Separation Distances for Portable and Mobile RF Communications Equipment
IEC 60601–1–2
Table 3-12:
Frequency of Transmitter
150 KHz to 26 MHz
26 MHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Equation
d = [3.5 / V
2
]√P
d = [3.5 / V
2
]√P
d = [3.5 / E
2
]√P
d = [3.5 / E
4
]√P
Rated Power of Transmit‐
ter (watts)
Distance (meters)
Distance (meters)
Distance (meters)
Distance (meters)
10 mW
1.2
1.2
1.2
2.3
100 mW
3.8
3.8
3.8
7.3
1
12
12
12
23 (*)
10
38
38
38
73
100
120
120
120
230
For transmitters rated at a power not listed above, the DISTANCE can be estimated using the
equation in the corresponding column, where P is the power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE: These are guidelines. Actual conditions may vary.
2.4.4 Limitations Management
Adhering to the distance separation recommended in Table 3-12, between 150 KHz & 2.5 GHz,
will reduce disturbances recorded at the image level, but may not eliminate all disturbances.
However, when installed and operated as specified herein, the system will maintain its essential
performance by continuing to acquire, display, and store diagnostic quality images safely.
For example, a 1W mobile phone (800 MHz to 2.5 GHz carrier frequency) shall be put 23 meters
(see (*) Table 3-12) apart from the Innova system (in order to avoid images interferences risks).
2.4.5 Use Limitation
The use of accessories, transducers, and cables other than those specified may result in degraded
ELECTROMAGNETIC COMPATIBILITY of the Innova system
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
Chapter 3 Physical Requirements of Room 63