2.4 Injector Mounting Requirements
Table accessory rail load consideration:
The maximum load per table accessory rail is 40 kg (88 lbs) at 150 mm (0.49 ft) (60
N.m or 44.25 ft/lbs). Therefore:
•
Only light extra load not exceeding 5 kg (11 lbs) at 100 mm (0.33 ft) (i.e IV pole
with its accessories, pressure head...) is authorized on the same table
accessory rail as the injector.
•
Never install injector and radiation protection on the same table accessory rail.
•
Typical installation on the front table accessory rail is Smart handle or Smart
box, Table Side System Control (TSSC), InnovaCentral/Touchscreen, Table
panning device and cables support.
•
If needed an optional rail can be installed at table foot end of the Omega V table
for other options.
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
164 2 Mounting Requirements