EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ V2

GE LOGIQ V2 User Guide

GE LOGIQ V2
381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #143 background imageLoading...
Page #143 background image
Measurement and Analysis
LOGIQ V2/LOGIQ V1 – User Guide 2-51
Direction 5610736-100 Rev. 9
Circumference and area (trace) measurement
Trace
To trace the circumference of a portion of the anatomy and
calculate its area:
1. Press Measure key to enter into measure mode, if
necessary.
Adjust Caliper Change in the Primary Menu to Trace.
NOTE: The available values for Caliper Change menu can be set
in Utility -> Measure -> Advanced preset -> Measure Key
Sequence.
2. To position the caliper at the start point, move the Trackball.
3. To fix the trace start point, press Set. The caliper changes to
an active caliper.
4. To trace the measurement area, move the Trackball around
the anatomy. A dotted line shows the traced area.
5. To complete the measurement, press Set. The system
displays the circumference and the area in the Results
Window.

Table of Contents

Other manuals for GE LOGIQ V2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V2 and is the answer not in the manual?

GE LOGIQ V2 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V2
CategoryMedical Equipment
LanguageEnglish

Related product manuals