System Start-Up
LOGIQ V2/LOGIQ V1User Guide 1-53
Direction 5610736-100 Rev. 9
Check System Date and Time
A warning message “Please check the system date and time are
correct” appears on the screen when the system is powered up.
This warning message appears for the possible reasons:
The system is not boot up for more than 14 days.
The system time has been changed by 24 hours earlier than
the current system time of last boot-up.
This warning message is to remind the user to check the system
date in case the system date and time is incorrect.
Figure 1-34. Check system date and time message
Move the cursor to OK and press Cursor key on the control
panel to select OK. The system enters scanning mode.
Check the system date and time. If it is incorrect, follow below
steps to reset the system date and time.
1. Enter Utility -> System -> General -> Date/Time.
2. Reset the system date and time.
3. Select Apply and then select OK.
4. Select Save.