Probe Overview
LOGIQ V2/LOGIQ V1 – User Guide 5-9
Direction 5610736-100 Rev. 9
Endocavitary Probe Handling Precautions
If the sterilization solution comes out of the endocavitary probe,
please follow the cautions below.
Sterile/sanitary sheaths are to be used on the probe during its
actual use with patients. Wearing gloves protects the patient
and operator.
Sterilant Exposure to Patient (e.g., Cidex)—Contact with a
sterilant to the patient’s skin or mucous membrane may cause
an inflammation. If this happens, refer to the sterilant’s
instruction manual.
Sterilant Exposure from Probe Handle/Connector to
Patient (e.g., Cidex)—DO NOT allow the sterilant to contact
the patient. Only immerse the probe to its specified level.
Ensure that no solution has entered the probe’s handle/
connector before scanning the patient. If sterilant comes into
contact with the patient, refer to the sterilant’s instruction
manual.
Endocavitary Probe Point of Contact—Refer to the
sterilant’s instruction manual.