System Care and Maintenance
LOGIQ V2/LOGIQ V1 – User Guide 3-41
Direction 5610736-100 Rev. 9
Trackball
1. Rotate the retainer counterclockwise until it can be removed
from the keyboard.
Figure 3-27. Rotate the retainer
2. Separate the trackball and the retainer. Wipe off any oil or
dust from the trackball, retainer and the trackball housing
using a cleaner or cotton swab.
3. Assemble the trackball and retainer, then put it into the
housing and rotate it clockwise until its notches are set in
position.
Other Peripheral Maintenance
Refer to the peripheral manufacture’s manuals for more
information.
CAUTION
When cleaning, make sure not to spill or spray any liquid into
the trackball housing (keyboard or system).