Performing an Exam
2-56 LOGIQ V2/LOGIQ V1 – User Guide
Direction 5610736-100
Rev. 9
Doppler Mode Measurements
Basic measurements below can be made in Doppler Mode.
• Velocity
• TAMAX and TAMEAN (Manual or Auto Trace)
• Two Velocities with the Time Interval and Acceleration
between them
• Time Interval
• Volume Flow
NOTE: The following instructions assume that you do the following:
1. In the B-Mode part of the display, scan the anatomy you
want to measure.
2. Go to the Doppler Mode part of the display.
3. Press Freeze.
Velocity
To measure velocity:
1. Press Measure key to enter into measure mode, if
necessary.
Adjust Caliper Change in the Primary Menu to Velocity.
An active caliper with a vertical and horizontal dotted line
displays.
2. To position the caliper at the desired measurement point,
move the Trackball.
3. To complete the measurement, press Set. The system
displays the velocity measurement in the Results Window.