Measurement and Analysis
LOGIQ V2/LOGIQ V1 – User Guide 2-55
Direction 5610736-100 Rev. 9
Intensity (Echo level) measurement
To make an echo level measurement:
1. Press Measure key to enter into measure mode, if
necessary.
2. Adjust Caliper Change in the Primary Menu to Intensity.
NOTE: The available values for Caliper Change menu can be set
in Utility -> Measure -> Advanced preset.
3. To position the caliper at the start point, move the Trackball.
4. To fix the trace start point, press Set. The caliper changes to
an active caliper.
5. To trace the measurement area, move the Trackball around
the anatomy. A dotted line shows the traced area.
6. To complete the measurement, press Set. The system
displays the echo level, as EL __ dB, in the Results Window.
NOTE: The echo level measurement is only available on a frozen
image, not on a B-paused image.