MEASure Subsystem
9 MEASure Subsystem
Measure commands measure the input voltage or current. They trigger the ac-
quisition of new data before returning the reading. Measurements are per-
formed by digitizing the instantaneous input voltage or current for a specified
measurement time, storing the results in a buffer, and calculating the value for
the specified measurement type.
MEASure[:SCALar]:CURRent[:DC]?
Initiates, triggers, and returns the averaged current measurement. Values re-
turned in amperes.
Group
MEASure
Syntax
MEASure[:SCALar]:CURRent[:DC]?
Arguments
None
Default Value
None
Returns
<NRf+>
Example
MEASure[:SCALar]:CURRent[:DC]?
Also see
None
Copyright © Itech Electronic Co., Ltd.
103