EasyManuals Logo
Home>ITech>Test Equipment>IT8000 Series

ITech IT8000 Series Programming Guide

ITech IT8000 Series
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #145 background imageLoading...
Page #145 background image
SOURce Subsystem
Syntax
[SOURce:]VOLTage:SLEW[:BOTH] <NRf+>
Arguments
<NRf+>
MINimum|MAXimum|DEFault|<value>
Setting range: MIN~MAX
Default Value
None
Example
VOLT:SLEW[:BOTH] 1
Query Syntax
[SOURce:]VOLTage:SLEW[:BOTH]?
Returns
<NRf+>
12.13 [SOURce:]VOLTage:SLEW:POSitive <NRf
+>
This command is used to set the voltage rise time.
Group
SOURce
Syntax
[SOURce:]VOLTage:SLEW:POSitive <NRf+>
Arguments
<NRf+>
MINimum|MAXimum|DEFault|<value>
Copyright © Itech Electronic Co., Ltd.
132

Table of Contents

Other manuals for ITech IT8000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8000 Series and is the answer not in the manual?

ITech IT8000 Series Specifications

General IconGeneral
BrandITech
ModelIT8000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals