INPut Subsystem
13.2 [SOURce:]INPut:DELay:FALL <NRf+>
This command is used to set the delay time for the input to be turned off.
Group
INPut
Syntax
[SOURce:]INPut:DELay:FALL <NRf+>
Arguments
<NRf+>
MINimum|MAXimum|DEFault|<value>
Setting range: MIN to MAX
Default Value
DEFault: 0
Example
INP:DEL:FALL 6
Query Syntax
[SOURce:]INPut:DELay:FALL?
Returns
<NRf+>
13.3 [SOURce:]INPut:DELay:RISE <NRf+>
This command is used to set the delay time for the input to be turned on.
Group
INPut