STATus Subsystem
ARB-Active 4 16
ARB has been trig-
gered and is being
executed
DLOG-
Active
5 32
DLOG has been trig-
gered and is being
executed
OFF 6 64
The [on/off] of the in-
strument is off
CC 7 128 Constant Current
CV 8 256
Constant Voltage
CW 9 512 Constant Power
STATus:QUEStionable[:EVENt]?
Queries the event register for the Questionable Status group. This is a read-only
register, which stores (latches) all events that are passed by the Operation NTR
and/or PTR filter. Reading the Questionable Status Event register clears it.
The value returned is the binary-weighted sum of all enabled bits in the
register. For example, with bit 2 (value 4) and bit 4 (value 16) set, the query
returns +20.
*RST has no effect on this register.
Group
STATus
Syntax
STATus:QUEStionable[:EVENt]?
Arguments
None
Default Value
None