STATus Subsystem
Query Syntax
STATus:QUEStionable:ENABle?
Returns
<NR1>
STATus:QUEStionable:PTRansition <NR1>
Sets the value of the PTR (Positive-Transition) registers. These registers serve
as a polarity filter between the Questionable Condition and Questionable Event
registers. When a bit in the PTR register is set to 1, then a 0-to-1 transition of
the corresponding bit in the Questionable Condition register causes that bit in
the Questionable Event register to be set. STATus:PRESet sets all bits in the
PTR registers and clears all bits in the NTR registers.
Group
STATus
Syntax
STATus:QUEStionable:PTRansition <NR1>
Arguments
<NR1>
A decimal value corresponding to the binary weighted sum of the register's bits.
Setting range: 0 to 65535.
Default Value
0
Example
Enable bit 3 and 4 in the questionable PTR register: STATus:QUEStionable:
PTRansition 24
Query Syntax
STATus:QUEStionable:PTRansition?
Copyright © Itech Electronic Co., Ltd.
78

Table of Contents

Other manuals for ITech IT8000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8000 Series and is the answer not in the manual?

ITech IT8000 Series Specifications

General IconGeneral
BrandITech
ModelIT8000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals