IEEE-488 Common Commands
*ESE <NR1>
Event status enable command. Sets the value in the enable register for the
Standard Event Status group. Each set bit of the register enables a correspond-
ing event. All enabled events are logically ORed into the ESB bit of the status
byte.
Any or all conditions can be reported to the ESB bit through the enable register.
To set the enable register mask, write a decimal value to the register using
*ESE.
*CLS does not clear the enable register, but does clear the event register.
Group
None
Syntax
*ESE <NR1>
Arguments
<NR1>
A decimal value corresponding to the binary-weighted sum of the register's bits.
The setting range is from 0 to 255.
Default Value
0
Returns
None
Example
Enable bits 3 and 4 in the enable register: *ESE 24
Also see
*ESE?