EasyManuals Logo
Home>Keithley>Measuring Instruments>6220

Keithley 6220 User Manual

Keithley 6220
454 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #101 background imageLoading...
Page #101 background image
Model 6220/6221 Reference Manual Sweeps 4-5
Return to Section 4 topics
Logarithmic staircase sweeps
This sweep is similar to the linear staircase sweep. The steps, however, are done
on a logarithmic scale as shown in the example sweep in Figure 4-3. This exam-
ple is a 5-point log sweep from 1mA to 10mA. The sweep delay parameter deter-
mines the time period for each step. As with the linear staircase sweep, the sweep
delay period is the same for all steps.
Figure 4-3
Logarithmic staircase sweep (example 5-point sweep from 1mA to 10mA)
Delay
Delay
Delay
Delay
Start
Bias
Stop
(10)
Delay
Log Points = 5
1
10
5.6234
3.1623
1.7783
Current
(mA)
Log
Scale
Sweep delay is the
same for each step.
0
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6220 and is the answer not in the manual?

Keithley 6220 Specifications

General IconGeneral
BrandKeithley
Model6220
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals