Model 6220/6221 Reference Manual SCPI Reference Tables 14-17
Return to Section 14 topics
Table 14-10
Trigger command summary
Command* Description Default Sec SCPI
INITiate Trigger initiation: 8
[:IMMediate] Initiate one trigger cycle. OFF
ABORt Reset trigger system. 8
ARM Arm layer commands: 8
[:SEQuence[1]]
[:LAYer[1]]
:SOURce <name> Select event detector: IMMediate, TIMer, BUS,
TLINk, BSTest, PSTest, NSTest, or MANual.
:SOURce? Query selected event detector.
:SIGNal Bypass ARM control source.
:TIMer <n> Set timer interval (seconds): 0 to 99999.99.
1msec resolution.
0.1
:TIMer? Query timer interval.
[:TCONfigure]
:DIRection <name> Control ARM source bypass: SOURce or
ACCeptor.
ACC
:DIRection? Query ARM source bypass.
[:ASYNchronous]
:ILINe <NRf> Set ARM input signal line: 1 to 6. 1
:ILINe? Query ARM input signal line.
:OLINe <NRf> Set ARM output signal line: 1 to 6. 2
:OLINe? Query ARM output signal line.
:OUTPut <name> Enable/disable ARM output trigger: TENTer,
TEXit or NONE.
NONE
:OUTPut? Query ARM output trigger.
TRIGger Trigger layer commands: 8
[:SEQuence[1]]
:SOURce <name> Select event detector: IMMediate or TLINk. IMM
:SOURce? Query selected event detector.
:SIGNal Bypass TRIG control source.
[:TCONfigure]
:DIRection <name> Control TRIG source bypass: SOURce or
ACCeptor.
ACC
:DIRection? Query TRIG source bypass.
[:ASYNchronous]
:ILINe <NRf> Set TRIG input signal line: 1 to 6. 1
* Arm layer commands control entire sweep. Trigger Layer commands control individual sweep steps.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com