8-2 Triggering Model 6220/6221 Reference Manual
Return to Section 8 topics
Trigger models
The trigger models control when to switch to the next point in a sweep and how
many sweeps to perform (see Section 4 for details on sweeps). Note that the trig-
ger configuration does not affect other Model 622x operation including normal out-
put currents and Model 6221 wave functions.
Front panel trigger model
The front panel trigger model is shown in Figure 8-1.
Figure 8-1
Trigger model for front panel operation
Arm-In
Source
Immediate
GPIB
Timer
Manual
TLink
ßSTest
ÝSTest
ßÝSTest
Arm Event
Detector
Trigger Event
Detector
Device
Action
No
Ye s
Idle
Another
Arm
?
Another
Trigger
?
Ye s
No
Arm
Layer
Trigger
Layer
Trigger-In
Source
Ye s
Halt
?
No
Once
Bypass
Arm Event
Detector
?
Never
Bypass
Trigger Event
Detector
?
Once
Never
Output Trigger
Source On/Off
Output
Trigger
(TL Exit)
On/Off
Sweep
Count
1
Immediate
TLINk
Press EXIT
Trigger 622x
Factory Default
Output Trigger
+
+
+
+
+
Sweep Parameters
Determine Number
of Steps
+
+
+
Sweep Step Occurs
During Device Action
(Controls
Sweep
Steps)
(Controls
Entire
Sweep)
Delay On/Off
+
On/Off
+
Output
Trigger (TL Enter)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com