Model 6220/6221 Reference Manual Delta, Pulse Delta, and Differential Conductance 5-31
Return to Section 5 topics
1. Use the :NVPResent command to determine if a suitable Model 2182/2182A with the
correct firmware revision is properly connected to the RS-232 port. This query com-
mand can be used for the system configuration shown in
Figure 5-2B on page 5-5.
2. See “Configuration settings,” on page 5-24 for details on these settings.
3. After setting up Delta using the above setup commands, the :ARM command arms
Delta. During the arming process, the Model 622x communicates with the Model
2182/2182A. Details on the “Arming process” are provided on page 5-24.
When armed, Delta will start when the Model 6221 is triggered. Delta can be “un-
armed” by sending the following command: SOURce:SWEep:ABORt.
The query form for the arm command (SOUR:DELT:ARM?) is used determine if Delta
is armed. A returned “1” indicates that Delta is armed. A “0” indicates that Delta is not
armed.
If the Model 6221 is already armed for a another action (e.g., Differential Conductance or
Pulse Delta), the Delta arm command will “un-arm” the other action and arm Delta.
Table 5-3
Delta commands
Command Description Default
[SOURce[1]]:DELTa:NVPResent? Queries connection to 2182A.
1
1 = yes, 0 = no
[SOURce[1]]:DELTa:HIGH <NRf> Sets high source value (amps).
2
<NRf> = 0 to 105e-3
1e-3
[SOURce[1]]:DELTa:LOW <NRf> Sets low source value (amps).
2
<NRf> = 0 to -105e-3
-1e3
[SOURce[1]]:DELTa:DELay <NRf> Sets Delta delay (seconds).
2
<NRf> = 0 to 9999.999 or INFinity
0
[SOURce[1]]:DELTa:COUNt <NRf> Sets the number of cycles to run.
2
<NRf> = 1 to 65536 (finite count) or INFinity
INF
[SOURce[1]]:SWEep:COUNt <NRf> Sets the number of measurement sets to
repeat.
2
<NRf> = 1 to 65536 (finite count) or INFinity
1
[SOURce[1]]:DELTa:CSWitch <b> Enable cold switching mode.
2
<b> = 0 or OFF, 1 or ON
0
[SOURCE[1]]:DELTa:CABort <b> Enable or disable Compliance Abort.
2
<b> = 0 or OFF, 1 or ON
0
[SOURce[1]]:DELTa:ARM Arms Delta.
3
[SOURce[1]]:DELTa:ARM? Queries Delta arm.
1 = armed, 0 = not armed.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6220 and is the answer not in the manual?

Keithley 6220 Specifications

General IconGeneral
BrandKeithley
Model6220
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals