EasyManuals Logo
Home>Keithley>Measuring Instruments>6220

Keithley 6220 User Manual

Keithley 6220
454 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #226 background imageLoading...
Page #226 background image
7-22 Wave Functions (6221 Only) Model 6220/6221 Reference Manual
Return to Section 7 topics
Programming sine waves
1. Restore defaults with this command:
*RST
2. Configure the waveform.
Examples – The following commands configure 1kHz sine wave with an
amplitude of 10mA, 1mA offset, and phase marker off:
3. Arm and trigger the waveform, turn on output:
4. To stop generating the waveform and turn the source output off before the
duration elapses, send this command:
SOUR:WAVE:FUNC SIN ‘ Select sine wave.
SOUR:WAVE:FREQ 1e3 ‘ Set frequency to 1kHz.
SOUR:WAVE:AMPL 1e-2 ‘ Set amplitude to 10mA.
SOUR:WAVE:OFFS 1e-3 ‘ Set offset to 1mA.
SOUR:WAVE:PMAR:STAT OFF ‘ Turn off phase marker.
SOUR:WAVE:DUR:TIME 5 ‘ 5 second duration.
SOUR:WAVE:RANG BEST ‘ Select best fixed source range.
SOUR:WAVE:ARM ‘Arm waveform.
SOUR:WAVE:INIT ‘ Turn on output, trigger waveform.
SOUR:WAVE:ABOR ‘ Stop generating waveform.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6220 and is the answer not in the manual?

Keithley 6220 Specifications

General IconGeneral
BrandKeithley
Model6220
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals