5-2 Delta, Pulse Delta, and Differential Conductance Model 6220/6221 Reference Manual
Return to Section 5 topics
Part 1
Overview
Keithley instrumentation requirements” for Delta, Pulse Delta, and Differential
Conductance are explained on page 5-4.
Section overview
This section is divided into two parts:
The topics in Part 1 provide support information for all three delta tests.
The topics in Part 2 (page 5-20) provide the details for each delta test.
Operation overview
The Model 6220 or 6221 Current Source can be used with a Model 2182/2182A
Nanovoltmeter to perform Delta and Differential Conductance. The Model 6221/
2182A combination can also perform Pulse Delta. These operations use a delta
current-reversal technique to cancel the effects of thermal EMFs.
The Model 622x provides a bipolar output current and the Model 2182/2182A per-
forms A/D conversions (measurements) at source high and source low points. An
averaging algorithm is then used to calculate the delta reading.
Delta – The Model 622x provides a square wave current output, and the Model
2182/2182A performs A/D conversions (measurements) at each high and low out-
put level. A 3-point moving-average algorithm is used to calculate Delta readings.
As shown in Figure 5-1A, the first three Model 2182/2182A A/D conversions (mea-
surements) yields the first Delta reading. Each subsequent Model 2182/2182A A/D
conversion then yields a single Delta reading. Every Delta reading uses the three
previous A/Ds to calculate Delta.
Pulse Delta – The Model 6221 outputs pulses and uses 3-point repeating-
average measurements to calculate Pulse Delta voltage. For each pulse, the
Model 2182A performs an A/D conversion (measurement) at pulse low, pulse
high, and pulse low. Each set of three A/D readings yield a single Pulse Delta
reading. Figure 5-1B shows Pulse Delta measurements. If device heating is a
concern, 2-point measurements can instead be used (2nd low pulse not
measured due to corruption from heat).
Differential Conductance – The Model 622x outputs a differential current (dI)
sweep and measures differential voltage (dV). This function uses a 3-point moving
average algorithm to calculate dV. With dI known and dV calculated, the Model
622x can then calculate differential conductance (dG) or differential resistance
(dR). Figure 5-1C shows differential conductance measurements.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6220 and is the answer not in the manual?

Keithley 6220 Specifications

General IconGeneral
BrandKeithley
Model6220
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals