2-20 Output Connections Model 6220/6221 Reference Manual
Return to Section 2 topics
Figure 2-17
Floating current source connections
Using a test fixture
Custom-built test fixture
A custom-built test fixture needs to a accommodate a variety of connection
requirements, including connections to an external voltmeter (such as the Keithley
Model 2182/2182A Nanovoltmeter).
Figure 2-18 shows a test fixture that can be custom-built. The included connectors
and terminals will accommodate any connection scheme covered in this section.
The internal wiring to the DUT will, of course, depend on the connection scheme
used.
High
Low
Output
High
Banana Plug Cable
Output
622x
622x
Test
circuit
LO
Test
circuit
NOTE Triax Output low must be disconnected from Earth Ground by using the following key-press
sequence:
Press TRIAX, select OUPUT LOW, select FLOATING.
The following command will also disconnect Output Low from Earth Ground:
OUTPut:LTEarth OFF
(red*)
(black*)
(red*)
* Boot color for alligator clip if
using supplied triax cable.
A) Inner shield connected to Output Low B) Inner shield connected to Cable Guard
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6220 and is the answer not in the manual?

Keithley 6220 Specifications

General IconGeneral
BrandKeithley
Model6220
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals