Model 6220/6221 Reference Manual Getting Started 1-17
Return to Section 1 topics
Perform the following steps to control the beeper:
1. Press MENU to display the MAIN MENU.
2. Using the “Menu navigation” keys (see page 1-20), place the cursor on
BEEPER and press ENTER.
3. Place the cursor on DISABLE or ENABLE and press ENTER. Beeper
enabled is the default setting.
4. Use the EXIT key to back out of the menu structure.
Remote programming – The following command controls the beeper:
Keyclick
With keyclick enabled, a “click” sound will be issued when a front panel key is
pressed. For the Model 6221, a “click” will also be issued when the rotary knob is
turned or pressed. Keyclick can only be controlled using remote operation:
Source preset
The PRES key can be used to set the source to a preset value. When the PRES
key is pressed, the source will set to the preset value (“PRES” message dis-
played). When the PRES key is pressed again, the unit will return to the original
source value. See “Source preset,” on page 3-13 for details on setting the preset
value.
Disabling the front panel
The front panel can be disabled. While disabled, display characters and annuncia-
tors are turned off. Also, all front panel controls (except LOCAL and DISP) are
locked out.
SYSTem:BEEPer:STATe <b> ‘
‘
‘
Enable or disable the beeper.
<b> = ON or OFF.
Default setting is beeper ON.
SYSTem:KCLick <b> ‘
‘
‘
Enable or disable keyclick.
<b> = ON or OFF.
Default setting is keyclick ON.
NOTE A single decimal point will be left on at the bottom right to
indicate that the unit is powered on.