4-10 Sweeps Model 6220/6221 Reference Manual
Return to Section 4 topics
Front panel sweep operation
Using the sweep configuration menu
To configure sweeps, press CONFIG then SWP, and then make your selections
from Table 4-2 below. See the detailed procedures for each sweep type in the fol-
lowing paragraphs.
NOTE User setups cannot be saved or recalled while a
sweep is armed or running. Attempting to do so
will generate error +413 Not allowed with mode
armed.
Table 4-2
Sweep configuration menu
Menu selection Description
TYPE
STAIR
LOG
CUSTOM
Select sweep type:
Set START, STOP, STEP, DELAY.
Set START, STOP, NO OF POINTS, DELAY.
Set #-POINTS, ADJUST POINTS, AUTO COPY.*
SWEEP-COUNT
FINITE
INFINITE
Choose sweep count:
Enter desired # of sweeps.
Continuously repeating sweeps.
SOURCE-RANGING
BEST
AUTO
FIXED
Select source ranging:
Use best range based on maximum sweep step.
Auto range based on individual sweep step.
Always stay on fixed source range.
COMPLIANCE-ABORT
NO
YES
Select compliance abort mode:
Do not abort sweep if compliance reached.
Abort sweep if compliance reached.
* Select AUTO COPY ON to automatically copy delay and compliance values to all sweep points.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6220 and is the answer not in the manual?

Keithley 6220 Specifications

General IconGeneral
BrandKeithley
Model6220
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals