Model 6220/6221 Reference Manual Status Structure 11-15
Return to Section 11 topics
• Bit B2, Over Temperature (Temp) — Set bit indicates that the internal
temperature limit has been exceeded. Check that the cooling vents are
clear and the heat sink is free of dust and dirt.
Error/status code: +112
• Bit B3, Compliance (Comp) — Set bit indicates that the Model 622x is in
compliance.
Error/status code: +114
• Bit B5, Reading Available (RAV) — Set bit indicates that a delta reading
has been taken and processed.
Error/status code: +306
• Bit B6, Trace Notify (TN) — Set bit indicates that the user-defined reading
number has been stored in the buffer (see TRACe:NOTify command in
Section 6).
Error/status code: +307
• Bit B7, Buffer Available (BAV) — Set bit indicates that there are at least
two readings stored in the buffer.
Error/status code: +308
• Bit B8, Buffer Half Full (BHF) — Set bit indicates that the buffer if half full.
Keep in mind that buffer size is set by the programmed number of sweep
points or delta measurement points (cycles).
Error/status code: +309
• Bit B9, Buffer Full (BFL) — Set bit indicates that the buffer is full. Keep in
mind that buffer size is set by the programmed number of sweep points or
delta measurement points (cycles).
Error/status code: +310
• Bit B12, Buffer Quarter Full (BQF) — Set bit indicates that the buffer is
quarter full. Keep in mind that buffer size is set by the programmed number
of sweep points or delta measurement points (cycles).
Error/status code: +312
• Bit B13, Buffer 3/4 Full (B3QF) — Set bit indicates that the buffer is 3/4
full. Keep in mind that buffer size is set by the programmed number of
sweep points or delta measurement points (cycles).
Error/status code: +313