14-6 SCPI Reference Tables Model 6220/6221 Reference Manual
Return to Section 14 topics
Table 14-3
Format command summary
Command Description Default
Ref
Sec SCPI
FORMat Reading format commands: 13
[:DATA] <type>[,<length>] Specify data format: ASCii, REAL,32, SREal,
DREal, or REAL,64
ASC
[:DATA]? Query data format.
:ELEMents <item list> Specify data elements: READing, TSTamp,
UNITs, RNUMber, SOURce, COMPliance,
AVOLtage (or ALL or DEFault)
READ,
TST
:ELEMents? Query data format elements
:BORDer <name> Specify byte order: NORMal or SWAPped. (Note)
:BORDer? Query byte order.
:SREGister <name> Select data format for reading status registers:
ASCii, HEXadecimal, OCTal, or BINary.
ASCii 11
:SREGister? Query format for reading status registers.
Note: *RST default is NORMal. SYSTem:PRESet default is SWAPped.
Table 14-4
Output command summary
Command Description Default Sec SCPI
OUTPut[1] Source output control commands:
[:STATe] <b> Turn output on or off (standby). OFF 3
[:STATe]? Query state (on or off) of output.
:LTEarth <b> Connect output low to earth ground (ON) or float
output low (OFF).
OFF 2
:LTEarth? Query output low connection.
:ISHield <name> Connect triax inner shield to OLOW (output low)
or cable GUARd.
OLOW 2
:ISHield? Query triax inner shield connection.
:RESPonse <name> Set the output response for 6221: FAST or SLOW. FAST 2
:RESPonse? Query output response.
:INTerlock Interlock: 2
:TRIPped? Returns a “0” if interlock is tripped (open) or a
“1”
if interlock is closed (output enabled).