14-8 SCPI Reference Tables Model 6220/6221 Reference Manual
Return to Section 14 topics
Table 14-6
Source command summary
Command Description Default Sec SCPI
SOURce[1] Current source output commands:
:CLEar Clearing the current source: 3
[:IMMediate] Set output to zero and then turn the output off.
:CURRent Current source:
[:LEVel] 3
[:IMMediate]
[:AMPLitude] <n> Set current source output (amps): -105e-3 to
105e-3.
0.0
:RANGe <n> Select a fixed source range: -105e-3 to 105e-3. 100e-3 3
:AUTO <b> Enable or disable source auto range. OFF
:AUTO? Query state of auto range.
:RANGe? Query present source range.
:COMPliance <NRf> Set voltage compliance (volts): 0.1 to 105 10 3
:COMPliance? Query voltage compliance.
:FILTer Analog filter: 3
[:STATe] <b> Enable or disable analog filter.
[:STATe]? Query state of analog filter. OFF
:STARt <n> Set start current (amps): -105e-3 to 105e-3 0 4
:STARt? Query start current.
:STOP <n> Set stop current (amps): -105e-3 to 105e-3 100e-3 4
:STOP? Query stop current.
:STEP <n> Set step current (amps): 1e-13 to 105e-3 1e-2 4
:STEP? Query step current.
:CENTer <n> Set center current (amps): -105e-3 to 105e-3 0 4
:CENTer? Query center current.
:SPAN <n> Set span current (amps): 2e-13 to 210e-3 100e-3 4
:SPAN? Query span current.
:DELay <n> Set source delay (seconds): 1e-3 to 999999.999 1 4
:DELay? Query source delay.
:SWEep Source sweep operation: 4
:SPACing <name> Select sweep type: LINear, LOGarithmic, or LIST LIN
:SPACing? Query sweep type.
:POINts <n> Set sweep points: 1 to 65535 11
:POINts? Query sweep points.
:RANGing <name> Select sweep ranging: AUTO, BEST, or FIXed. BEST
:RANGing? Query sweep ranging.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com