EasyManuals Logo
Home>Keithley>Switch>7001

Keithley 7001 User Manual

Keithley 7001
353 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #159 background imageLoading...
Page #159 background image
IEEE-488 Reference
5-40
5.10.2
*ESE
event status enable
To set the contents of the Standard Event Status Enable Register.
*ESE <NRf>
<NRf> = 0 Clear (reset) register
= 1 Set OPC (B0) of enable register
= 4 Set QYE (B2) of enable register
= 8 Set DDE (B3) of enable register
= 16 Set EXE (B4) of enable register
= 32 Set CME (B5) of enable register
= 64 Set URQ (B6) of enable register
= 128 Set PON (B7) of enable register
Power-up Clears register
*CLS No effect
*RST No effect
:STATus:PRESet No effect
DCL, SDC No effect
The *ESE command is used to set the contents of the Standard Event Status Enable Register.
This command is sent with the decimal equivalent of the binary value that determines the de-
sired state (0 or 1) of the bits in the register.
This register is used as a mask for the Standard Event Status Register. When a standard event is
masked, the occurrence of that event will not set the Event Summary Bit (ESB) in the Status
Byte Register. Conversely, when a standard event is unmasked (enabled), the occurrence of that
event will set the ESB bit. For information on the Standard Event Status Register and descrip-
tions of the standard event bits, see paragraph 5.10.4. The Status Byte Register is described in
paragraph 5.6.8.
A cleared bit (bit set to “0”) in the enable register will prevent (mask) the ESB bit in the Status
Byte Register from setting when the corresponding standard event occurs. A set bit (bit set to
“1”) in the enable register will allow (enable) the ESB bit to set when the corresponding standard
event occurs. For details on the status register structure, refer to paragraph 5.6.
The Standard Event Status Enable Register is shown in Figure 5-13. Notice that the decimal
weight of each bit is included in the illustration. The sum of the decimal weights of the bits that
you wish to be set is the value that is sent with the *ESE command. For example, to set the CME
and QYE bits of the Standard Event Status Enable Register, send the following command:
*ESE 36
Purpose
Format
Parameters
Defaults
Description
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Other manuals for Keithley 7001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 7001 and is the answer not in the manual?

Keithley 7001 Specifications

General IconGeneral
BrandKeithley
Model7001
CategorySwitch
LanguageEnglish

Related product manuals