IEEE-488 Reference
5-125
These action commands are used to bypass the specified layer of the Trigger Model. They are
used when you do not wish to wait for the programmed event. Note from the Trigger Model
(Figure 5-13) that :arm:lay2:imm and :trig:imm also bypass the Delay.
The instrument must be waiting for the appropriate event when the command is sent. Otherwise,
an error will occur and the command will be ignored. When the command is sent, the bypass
will occur and operation will continue on.
Note that :IMMediate is not an instrument setup command. As soon as this command is execut-
ed, it will attempt to bypass the specified trigger model layer.
OUTPUT 707; “:arm:imm” ! Bypass arm control source
:COUNt <n>
:ARM[:SEQuence[1]][:LAYer[1]]:COUNt <n> Set Arm Layer 1 count
:ARM[:SEQuence[1]]:LAYer2:COUNt <n> Set Arm Layer 2 count
:TRIGger[:SEQuence[1]]:COUNt <n> Set Trigger Layer count
<n> = 1 to 99999 Specify count
= INF Sets count to infinite
= DEFault Sets count to 1
= MINimum Sets count to 1
= MAXimum Sets count to 9999
:arm:coun <n>
:arm:lay2:coun <n>
:trig:coun <n>
Power-up Saved power-on setup
*RST 1 (all layers)
:SYSTem:PRESet 1 (Arm Layer 1)
INF (Arm Layer 2)
Scan-list-length (Trigger Layer)
:COUNt? Query programmed count
:COUNt? DEFault Query *RST default count
:COUNt? MINimum Query lowest allowable count
:COUNt? MAXimum Query largest allowable count
Short-form formats: :arm:coun? :arm:lay2:coun?
:arm:coun? xxx :arm:lay2:coun? xxx
:trig:coun?
:trig:coun? xxx
where; xxx = def, min or max
Description
Programming
example
5.18.4
Parameters
Short-form
formats
Defaults
Query
commands
Trigger subsystem
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Other manuals for Keithley 7001

Question and Answer IconNeed help?

Do you have a question about the Keithley 7001 and is the answer not in the manual?

Keithley 7001 Specifications

General IconGeneral
Max Switching Voltage300 V
Max Switching Current1 A
Operating Temperature0°C to 50°C
Switching Speed1 ms
InterfaceGPIB, RS-232
Switching TypeVarious, depending on installed switch cards (e.g., electromechanical, solid-state)

Summary

Warranty

Limitation of Warranty

Limits of warranty coverage, remedies, and exclusions.

Safety Precautions

7001 High Density Switch System Specifications

SYSTEM

Core system parameters including capacity, memory, timing, triggers, and I/O.

ANALOG BACKPLANE

Details on analog backplane signals, voltage, current, and isolation.

1 General Information

1.3 Warranty Information

Details and procedures related to product warranty.

1.5 Safety Symbols and Terms

Explanation of safety symbols and terms used throughout the manual.

1.6 Specifications

Technical specifications of the Model 7001 instrument.

2 Card Installation

2.2 Model 701X Series Card Installation

Steps for installing 701X series cards into the mainframe.

2.3 Non-701X Series Cards

Steps for installing non-701X series cards into the mainframe.

4 Front Panel Operation

4.2 Power-up Procedure

Steps for connecting power and starting the instrument.

5 IEEE-488 Reference

7 Maintenance

7.3 Static-Sensitive Devices

Precautions for handling static-sensitive components.

7.5 Disassembly

Procedures for removing instrument covers and boards.

7.6 Diagnostics

How to run the instrument's built-in diagnostic tests.

7.8 Troubleshooting

Guidance for diagnosing and fixing instrument malfunctions.

8 Replaceable Parts

8.5 How to Obtain Factory Service

Process for returning the instrument for factory service.

6 Theory of Operation

6.2 Overall Functional Description

General functional overview of the Model 7001.

6.3 Control Function

Description of the instrument's internal control mechanisms.

Related product manuals