IEEE-488 Reference
5-125
These action commands are used to bypass the specified layer of the Trigger Model. They are
used when you do not wish to wait for the programmed event. Note from the Trigger Model
(Figure 5-13) that :arm:lay2:imm and :trig:imm also bypass the Delay.
The instrument must be waiting for the appropriate event when the command is sent. Otherwise,
an error will occur and the command will be ignored. When the command is sent, the bypass
will occur and operation will continue on.
Note that :IMMediate is not an instrument setup command. As soon as this command is execut-
ed, it will attempt to bypass the specified trigger model layer.
OUTPUT 707; “:arm:imm” ! Bypass arm control source
:COUNt <n>
:ARM[:SEQuence[1]][:LAYer[1]]:COUNt <n> Set Arm Layer 1 count
:ARM[:SEQuence[1]]:LAYer2:COUNt <n> Set Arm Layer 2 count
:TRIGger[:SEQuence[1]]:COUNt <n> Set Trigger Layer count
<n> = 1 to 99999 Specify count
= INF Sets count to infinite
= DEFault Sets count to 1
= MINimum Sets count to 1
= MAXimum Sets count to 9999
:arm:coun <n>
:arm:lay2:coun <n>
:trig:coun <n>
Power-up Saved power-on setup
*RST 1 (all layers)
:SYSTem:PRESet 1 (Arm Layer 1)
INF (Arm Layer 2)
Scan-list-length (Trigger Layer)
:COUNt? Query programmed count
:COUNt? DEFault Query *RST default count
:COUNt? MINimum Query lowest allowable count
:COUNt? MAXimum Query largest allowable count
Short-form formats: :arm:coun? :arm:lay2:coun?
:arm:coun? xxx :arm:lay2:coun? xxx
:trig:coun?
:trig:coun? xxx
where; xxx = def, min or max
Description
Programming
example
5.18.4
Parameters
Short-form
formats
Defaults
Query
commands
Trigger subsystem