SCPI Conformance Information
F-5
Table F-2
Syntax of non-SCPI commands implemented by Model 7001
Command Description
:DISPlay
:SMESsage <b>
:SMESsage?
Enable (1 or ON) or disable (0 or OFF) displayed status messages.
Request state of status message display.
:OUTPut
:TTLn
:LSENse AHIGh|ALOW
:LSENse?
Path and command to specify output line ÒnÓ (1-4):
Set polarity; active high (AHIG) or active low (ALOW).
Request polarity of speciÞed output line (n).
[:ROUTe]
:SCAN?
:FCHannels <list>
:FCHannels?
:CONFigure
:SCHannel <b>
:SCHannel?
:CPAir <b>
:CPAir?
:SLOT1
:CTYPE <type>
:CTYPE?
:POLE <n>
:POLE?
:STIMe <num>
:STIMe?
:SLOT2
:CTYPE <type>
:CTYPE?
:POLE <n>
:POLE?
:STIMe <num>
:STIMe?
:MEMory
:SAVe M<num>
:RECall M<num>
Request the programmed scan list.
SpeciÞes channels that cannot be closed (forbidden).
Request list of restricted channels.
ConÞguration command path:
Enable (1 or ON) or disable (0 or OFF) Single Channel.
Request status of Single Channel mode.
Enable (1 or ON) or disable (0 or OFF) Channel Pairing.
Request status of Channel Pairing.
Path to conÞgure CARD 1:
Specify the card type installed in the slot.
Request the card type installed in the slot.
Select pole mode (1, 2 or 4).
Request the programmed number of poles.
Set relay settling time in seconds (0 to 99999.999).
Request the programmed relay settling time.
Path to conÞgure CARD 2:
Specify the card type installed in the slot.
Request the card type installed in the slot.
Select pole mode (1, 2 or 4).
Request the programmed number of poles.
Set relay settling time in seconds (0 to 99999.999).
Request the programmed relay settling time.
Path to program memory:
Saves channel pattern at memory location (1-100).
Recalls channel pattern from memory location (1-100).
:SENSe[1]
:TTL1
:DATA?
Command path for input port.
Read digital input port.
:SENSe2
:DATA? <list>
Command path:
Read speciÞed input channels for slot 1.
:SENSe3
:DATA? <list>
Command path:
Read speciÞed input channels for slot 2.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Other manuals for Keithley 7001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 7001 and is the answer not in the manual?

Keithley 7001 Specifications

General IconGeneral
BrandKeithley
Model7001
CategorySwitch
LanguageEnglish

Related product manuals