SCPI Conformance Information
F-4
Trigger commands
:TRIGger[:SEQuence[1]]
:IMMediate
:COUNt <n>|INFinite
:AUTo ON/OFF
:COUNt?
:DELay <num>
:DELay?
:SOURce HOLD|IMMediate|
MANual|BUS|EXTernal|TIMer
:SOURce?
:SIGNal
:TIMer <num>
:TIMer?
Path to program channel layer of scan:
Scan channels immediately. (Loop around control source.)
Program number of channels (1 to 9999, or INFinite).
Use scan-list-length.
Request programmed channel count.
Program delay (0 to 99999.999 sec)
Request programmed delay period.
Select Trigger Layer Control Source.
Request programmed source (event).
Bypass channel control source one time only.
Set timer interval (0 to 99999.999 sec).
Request the programmed timer interval.
Table F-1 (continued)
Syntax of SCPI confirmed commands implemented by Model 7001
Command Description