EasyManuals Logo
Home>Keithley>Switch>7001

Keithley 7001 User Manual

Keithley 7001
353 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #202 background imageLoading...
Page #202 background image
IEEE-488 Reference
5-83
[:ROUTe] subsystem
Power-up Saved power-on setup
*RST Off
:SYSTem:PRESet Off
:SCHannel? Query status of Single Channel
Short-form formats: :conf:sch?
Response message: 0 (off) or 1 (on)
The Single Channel mode (when enabled) is used to prevent more than one channel from clos-
ing at a time. When a single channel is closed, the previous closed channel will open. This en-
sures that even with random channel closures, only one channel can be closed at a time.
Toggling the Single Channel mode from off to on causes all channels to open. For complete de-
tails on Single Channel, see Section 4.
10 OUTPUT 707; “:conf:sch on; sch?”
20 ENTER 707; A$
30 PRINT A$
40 END
Line 10 Enables Single Channel and then queries its status.
Line 20 Addresses the 7001 to talk.
Line 30 Displays the status of Single Channel (1; on).
:CPAir <b>
[:ROUTe]:CONFigure:CPAir <b> Control Card Pair
<b> = 0 or OFF Disable Card Pair.
= 1 or ON Enable Card Pair.
:conf:cpa <b>
Power-up Saved power-on setup
*RST Disables Card Pair
:SYSTem:PRESet Disables Card Pair
:CPAir? Query status of Card Pair
Short-form format: :conf:cpa?
Response message: 0 (off) or 1 (on)
Defaults
Query
command
Description
Programming
example
5.13.7
Parameters
Short-form
format
Defaults
Query
command
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Other manuals for Keithley 7001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 7001 and is the answer not in the manual?

Keithley 7001 Specifications

General IconGeneral
BrandKeithley
Model7001
CategorySwitch
LanguageEnglish

Related product manuals