IEEE-488 Reference
5-123
NOTE
Front panel operation and IEEE-488 operation use different nomencalture for
the three layers of the Trigger Model and is summarized as follows:
Trigger Model Layer Nomenclature
Layer Name; Front Layer Name; IEEE-488
Layer Panel Operation Bus Operation
First layer Arm Layer Arm Layer 1
Second layer Scan Layer Arm Layer 2
Third layer Channel Layer Trigger Layer
:INITiate commands
[:IMMediate]
:INITiate[:IMMediate] Take 7001 out of idle
:init
This command takes the Model 7001 out of the idle state. After the completion of a scan, the
instrument will return to the idle state (if continuous initiation is disabled; see next command).
This is an overlapped command (see *OPC, *OPC?, *TRG and *WAI).`
OUTPUT 707; “:init” ! Takes 7001 out of the idle state.
:CONTinuous <b>
:INITiate:CONTinuous <b> Control continuous initiation
<b> = 0 or OFF Disable continuous initiation
= 1 or ON Enable continuous initiation
:init:cont <b>
Power-up Saved power-on setup
*RST OFF
:SYSTem:PRESet ON
5.18.1
Short-form
format
Description
Programming
example
Parameters
Short-form
format
Defaults
Trigger subsystem